Visual abstractsVisuals
ResearchJournal of Urology · Jan 21, 2026

First-In-Human Assessment of the UroActive Electronic Artificial Urinary Sphincter for Treating Male Patients with Stress Urinary Incontinence

Mozer P, Beaugerie A, Perrouin-Verbe MA, Denormandie A, Cotte J, Plassais C, et al
UrologyAdult
Read the full paper free at PubMed Central ↗

Abstract

Purpose To evaluate the feasibility, safety, and efficacy during the first year of use of the UroActive device in male patients with stress urinary incontinence. Materials and methods This prospective, multicenter, single-arm, first-in-human study enrolled 6 male patients with bothersome stress urinary incontinence unresponsive to conservative or surgical treatments. After UroActive implantation, follow-up was planned for 5 years (postactivation). Study end points included successful device activation, revision, and explantation rates. Other end points assessed 24 hours pad-weight test reduction, pad usage, uroflowmetry, postvoid residual urine, and patient-reported outcomes (International Consultation Incontinence Questionnaires UI Short Form, EQ-5D, I-QOL, Urinary Symptom Profile). Adverse events were collected throughout the study. Results All 6 patients underwent successful implantation and device activation (100%). No explants or revisions were required during the 365 days postactivation. At 365 days postactivation, the pad weight was reduced by ≥ 50% in all patients and reduced by ≥ 75% in 5 (83.3%). The median pad use decreased from 3.7 to 1.0 pad per day. Uroflowmetry remained within acceptable ranges, and postvoid residual urine was minimal. Quality of life scores significantly improved across all instruments. Nine device deficiencies were reported, all involving the patient remote control. No unanticipated serious adverse events occurred. Conclusions In men with stress urinary incontinence, the UroActive device demonstrates promising feasibility, safety, and functional outcomes during the 365 days postactivation. Patients experienced substantial symptom improvement and high satisfaction, with no surgical revisions. Larger studies with extended follow-up are needed to confirm these findings and assess long-term device performance.

Topics

Humans; Male; Urinary Incontinence, Stress; Urinary Sphincter, Artificial; Prospective Studies; Middle Aged; Aged; Feasibility Studies; Treatment Outcome; Prosthesis Implantation; Prosthesis Design; Follow-Up Studies

DOI: 10.1097/JU.0000000000004945 · PubMed: 41564274

Cite this article

Mozer P, Beaugerie A, Perrouin-Verbe MA, Denormandie A, Cotte J, Plassais C. First-In-Human Assessment of the UroActive Electronic Artificial Urinary Sphincter for Treating Male Patients with Stress Urinary Incontinence. J Urol. 2026. doi:10.1097/JU.0000000000004945

RIS (EndNote, Zotero, Mendeley)BibTeX

Spotted a mistake? Suggest a correction

Similar articles

J Urol · Jul 15, 2026 · Research · Urology, Surgery

Frailty Is Associated With Postoperative Complications and Device Removal and Revision in Men Undergoing Surgery for Stress Urinary Incontinence

JAMA · May 5, 2025 · Randomized trial · Urology Visual abstract

OnabotulinumtoxinA and midurethral sling showed similar results for treating mixed urinary incontinence

JAMA Netw Open · Apr 1, 2026 · Randomized trial · Urology Visual abstract

Pelvic floor workout during pregnancy reduced urinary leakage in women six weeks after childbirth

Eur Urol Oncol · Nov 22, 2025 · Research · Urology, Oncology

A Novel Transurethral Robotic Platform for Performing En Bloc Resection of Bladder Tumour: Results from the First Phase of the VIABLE Trial

BJU Int · Nov 27, 2025 · Clinical trial · Urology

The LaserSAFE technique for margin assessment during radical prostatectomy: a feasibility study