Image quality and lesion detectability in simulated [124I]MIBG cardiac imaging across 3-generation PET/CT scanners using a phantom model: Implications for activity and acquisition time reduction in clinical practice

Aghapour Zangeneh F, Zeimpekis K, Gnesin S, Knappe LM, Garibotto V, Kamani CH, et al
Radiology & ImagingAdult
Read the full paper free at PubMed Central ↗

Abstract

Background Cardiac sympathetic imaging with iodine-labeled metaiodobenzylguanidine (MIBG) is clinically relevant for assessing autonomic dysfunction and arrhythmic risk in heart failure. The development of positron-emitting 124I-MIBG and new-generation PET/CT systems, particularly long axial field-of-view (LAFOV) scanners, may improve image quality and regional defect assessment while enabling protocol optimization. This study compared simulated cardiac 124I-MIBG image quality and lesion detectability across three PET/CT generations, analog standard axial field-of-view (SAFOV), digital SAFOV, and LAFOV, and explored the potential for acquisition-time and activity reduction. Methods An anthropomorphic torso phantom with cardiac insert was prepared to simulate physiologic 124I-MIBG biodistribution. Two configurations were studied: a pathological model with a transmural defect (TD) and a non-transmural defect (NTD), and a normal model with homogeneous myocardial uptake. Acquisitions were performed on three PET/CT systems: analog SAFOV (Biograph mCT), digital SAFOV (Biograph Vision Edge 600), and LAFOV (Biograph Vision Quadra). List-mode data were acquired for 10 min and reconstructed to simulate 1-, 5-, and 10-min acquisitions, yielding nine datasets. Qualitative assessment included lesion detectability and overall image quality scored on 5-point Likert scales by two experienced readers. Quantitative analysis included signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR). Exploratory pairwise Wald-type comparisons were performed using propagated measurement uncertainty. Results LAFOV PET/CT showed the best qualitative and quantitative performance overall. Lesion detectability was highest on LAFOV at all acquisition times, with digital SAFOV intermediate and analog SAFOV lowest. Image quality likewise favored LAFOV, especially at 5 and 10 min. In the pathological phantom, SNR increased from 25.9 to 75.1 on LAFOV, from 18.7 to 41.8 on digital SAFOV, and from 15.5 to 17.1 on analog SAFOV. At 10 min, exploratory uncorrected pairwise comparisons showed nominally higher SNR on LAFOV than on analog SAFOV PET/CT (75.1 vs 17.1, p = 0.028). For the transmural defect, CNR also increased most on LAFOV, with a nominally higher value versus analog SAFOV at 10 min (67.6 vs 14.5, p = 0.044). By contrast, for the low-contrast non-transmural defect, CNR was numerically higher on LAFOV at longer acquisition times, but the absolute inter-scanner differences were smaller and did not reach statistical significance. Conclusion LAFOV PET/CT provides superior image quality and lesion detectability for simulated cardiac 124I-MIBG imaging and supports shorter acquisitions and potentially lower administered activity. However, the advantage was most pronounced for high-contrast transmural defects, whereas the benefit for low-contrast non-transmural defects was less robust. Reliable detection of subtle partial innervation defects remains challenging and requires dedicated clinical validation.

Topics

Positron Emission Tomography Computed Tomography; Phantoms, Imaging; 3-Iodobenzylguanidine; Heart; Humans; Time Factors; Image Processing, Computer-Assisted; Signal-To-Noise Ratio; Iodine Radioisotopes

DOI: 10.1007/s00259-026-08065-0 · PubMed: 42440083

Cite this article

Aghapour Zangeneh F, Zeimpekis K, Gnesin S, Knappe LM, Garibotto V, Kamani CH. Image quality and lesion detectability in simulated [124I]MIBG cardiac imaging across 3-generation PET/CT scanners using a phantom model: Implications for activity and acquisition time reduction in clinical practice. Eur J Nucl Med Mol Imaging. 2026. doi:10.1007/s00259-026-08065-0

RIS (EndNote, Zotero, Mendeley)BibTeX

Spotted a mistake? Suggest a correction

Similar articles

J Nucl Med · Feb 2, 2026 · Research · Radiology & Imaging

Development of a Prototype AWSM-PET Device for Augmented Whole-Body PET Imaging and Initial Human Study

J Nucl Med · Aug 3, 2026 · Research · Radiology & Imaging

Revisiting Perfusion in the Forgotten Right Ventricle: Artificial Intelligence-Enhanced Quantification on PET/CT

Eur Radiol · Jan 15, 2026 · Research · Radiology & Imaging

TomoRay cranial: synthesis of cranial CT imaging from biplanar radiographs using a generative adversarial network

Eur J Nucl Med Mol Imaging · Dec 16, 2025 · Research · Radiology & Imaging

No reinduction of clinically relevant radioiodine uptake after lenvatinib treatment in radioidine-refractory differentiated thyroid cancer

Eur J Nucl Med Mol Imaging · May 26, 2026 · Research · Radiology & Imaging

Diagnostic performance of [18F]Tetrafluoroborate PET versus 131I scintigraphy in differentiated thyroid cancer: a prospective study