Endotracheal tubes delivered more accurate ventilation than supraglottic airways in cadavers
Influence of different airway devices on intra-arrest ventilation during bag-valve-device ventilation - a prospective randomized controlled cadaver study
Papers we list that report this registered trial, oldest first. Read a secondary analysis alongside the main results.
- Main results
Tracheal intubation was associated with better ventilation during continuous resuscitation than supraglottic airway devices Resuscitation 2025-09 - Related analysis
This paper Crit Care 2025-12
Why it matters: The optimal ventilation strategy during cardiac arrest remains unclear, particularly regarding how different airway devices affect manual ventilation accuracy.
- Eleven cadavers were ventilated using endotracheal tubes or supraglottic airways to compare tidal volume accuracy and leakage.
- Endotracheal tubes showed a lower difference between ideal and expiratory tidal volume (142.2 ml vs 202.0 ml) than supraglottic airways.
- Supraglottic airways had higher leakage volume (39.0% vs 17.9%) and lower peak pressures than endotracheal tubes.
Headline and key points written with AI from the paper’s abstract, with every number checked against it. Read the paper before relying on them.
Was this summary helpful?
Subject terms
Headline, design, result, key points, caveat and citation as plain text.
Cite this article
Lohmann J, Brand-Saberi B, Ullrich M, et al. Influence of different airway devices on intra-arrest ventilation during bag-valve-device ventilation - a prospective randomized controlled cadaver study. Crit Care. 2025;29(1):519. doi:10.1186/s13054-025-05713-z
Spotted a mistake? Suggest a correction