Incorrect esophageal probe placement leads to lower and more variable core temperature readings in cold air
Influence of esophageal temperature probe tip placement on core temperature measurement accuracy in cold environments: a randomized crossover trial with implications for cardiac arrest management
Why it matters: The effect of proximal misplacement of esophageal probes on core temperature measurement accuracy during cold air exposure remains unknown.
- Fifteen healthy volunteers participated in a randomized crossover study comparing correct esophageal probe placement behind the heart to incorrect placement 5 cm above the tracheal bifurcation.
- During 20 min of exposure to -20 °C, mean core temperature was on average 0.6 °C lower when the probe was incorrectly positioned.
- The study also found that temperature fluctuations were greater when the esophageal probe was placed in the incorrect proximal position.
Headline and key points written with AI from the paper’s abstract, with every number checked against it. Read the paper before relying on them.
Was this summary helpful?
Cite this article
Roveri G, Dal Cappello T, Hofer A, Breidt F, Kofler O, Popp E. Influence of esophageal temperature probe tip placement on core temperature measurement accuracy in cold environments: a randomized crossover trial with implications for cardiac arrest management. Resusc Plus. 2025. doi:10.1016/j.resplu.2025.101087