ResearchEuropean Radiology · Jun 24, 2026
Lung nodule detection and potential impact on guideline-based management: a retrospective post-market evaluation of three commercial software systems
Jöbstl A, Luger AK, Nilica B, Kocher F, Sonnweber T, Tancevski I, et al
Abstract
Objectives
To evaluate three commercial AI software tools for pulmonary nodule detection and segmentation and to assess their impact on guideline-based management recommendations. Materials and methods
A total of 740 CT and PET-CT studies from clinical routine were analyzed using three software tools (S1, S2, S3). We compared the total number of detected nodules and "actionable" nodules (per British Thoracic Society (BTS) definition). We further evaluated how measurement variations between tools affected hypothetical management according to Fleischner Society and BTS guidelines for incidental nodules. Results
The tools differed significantly in the total number of detections (S1: 1336; S2: 1060; S3: 1536; p Conclusion
While the three included AI tools show comparable performance in detecting actionable nodules, minor measurement variations significantly impact downstream management when using guidelines with narrow thresholds, such as the BTS criteria. Fleischner Society guidelines appear more robust to these inter-software variations. Key points
Question How do commercial software tools for pulmonary nodule detection perform in real-world settings and impact hypothetical management under BTS and Fleischner guidelines? Findings Detection of actionable nodules was comparable across all tools, but small absolute measurement variations triggered significantly more 1-year follow-up recommendations under BTS guidelines. Clinical relevance AI software can cause inconsistent BTS-based management due to narrow thresholds, while Fleischner criteria appear more stable. Frequent detection of benign lesions potentially poses a risk of overdiagnosis and overtreatment in standalone AI-based reporting.
Topics
Humans; Software; Solitary Pulmonary Nodule; Retrospective Studies; Practice Guidelines as Topic; Tomography, X-Ray Computed; Lung Neoplasms; Positron Emission Tomography Computed Tomography
DOI: 10.1007/s00330-026-12702-5 · PubMed: 42343062
Spotted a mistake? Suggest a correction
More in Radiology & Imaging
ResearchInsights Imaging · Nov 20, 2026 · Radiology & Imaging
Enhanced prognosis stratification following surgery for pancreatic ductal adenocarcinoma to optimize clinical decisions by reflecting tumor heterogeneity and microenvironment
ResearchInsights Imaging · Nov 19, 2026 · Radiology & Imaging
Diagnostic performance of dual-phase photon-counting detector for localization of parathyroid adenomas after inconclusive first-line imaging
ResearchInsights Imaging · Nov 18, 2026 · Radiology & Imaging
ResearchInsights Imaging · Nov 18, 2026 · Radiology & Imaging
Classic signs in head and neck imaging-a quick guide for the radiology resident
ResearchInsights Imaging · Nov 15, 2026 · Radiology & Imaging
Carotid CTA with automated plaque quantification for risk stratification of coexisting obstructive coronary artery disease