Visual abstractsVisuals
ResearchEuropean Radiology · Dec 22, 2025

Neonatal cerebral imaging: feasibility and image quality using an MR-compatible incubator equipped with a 16-channel head coil

Tischendorf P, Wiesmann M, Omran H, Krähling T, Sandkötter J, Heindel W, et al
Radiology & ImagingPediatric

Study type: Observational study: it can show a link, but not prove that one thing causes another.

Read the full paper free at PubMed Central ↗

Abstract

Objectives To address the needs of preterm and term neonates undergoing cerebral magnetic resonance imaging (cMRI), using a recently developed MR-compatible incubator (icMRI) equipped with a 16-channel receiver head coil was evaluated for feasibility and image quality compared to conventional cMRI (ccMRI) using a 16-channel adult head coil. Materials and methods One hundred three consecutive neonates (46 females; 40 ± 3 gestational weeks) underwent clinically indicated cMRI, 53 were examined with an incubator, and 50 without. Normally distributed variables of the two groups were compared using the Student's t-test. The Wilcoxon signed-rank test was applied for unequal variances. Image quality was subjectively rated on a 5-point scale by two pediatric radiologists, with interobserver agreement and signal-to-noise ratios in the parieto-occipital white matter (SNRc) and basal ganglia (SNRb) calculated. Interobserver agreement was assessed using the quadratic weighted Cohen's kappa test (ϰ). Mean SNR values between groups were compared using the Student's t-test. Results Neonates who underwent icMRI had a significantly lower weight at imaging (p = 0.04). Total imaging time was significantly shorter (p = 0.03) using an incubator. icMRI led to a significant increase in SNRc and SNRb across all sequences (p ≤ 0.006), while diagnostic image quality improved from 3.5 to 4.3 points, and image motion artifacts decreased from 31% to 14% with substantial to almost perfect interobserver agreement. Furthermore, icMRI was associated with reduced need for general anesthesia (p = 0.03). Conclusion icMRI is correlated with better feasibility, improved image quality, shorter scan times, and reduced need for general anesthesia compared to ccMRI. Key points QuestionNeonatal cerebral MRI requires a precisely controlled environment, including optimal temperature, humidity, and real-time vital monitoring to ensure safety and stability during imaging. Findings A recently developed icMRI with an integrated 16-channel radiofrequency receiving head coil is associated with improved feasibility and image quality in neonatal cMRI. Clinical relevance Newborns benefit from a recently developed icMRI with an integrated 16-channel head coil that optimizes safety, handling, monitoring, and transport, while also correlating with improved image quality, reduced scan time's and reduced need for general anesthesia.

Topics

Humans; Infant, Newborn; Magnetic Resonance Imaging; Female; Feasibility Studies; Male; Signal-To-Noise Ratio; Incubators, Infant; Brain; Equipment Design

DOI: 10.1007/s00330-025-12219-3 · PubMed: 41428043

Spotted a mistake? Suggest a correction

More in Radiology & Imaging

Insights Imaging · Nov 20, 2026 · Research · Radiology & Imaging

Integrated prognostic system predicts disease-free survival in pancreatic cancer patients

Insights Imaging · Nov 19, 2026 · Research · Radiology & Imaging

Dual-phase photon-counting CT localized parathyroid adenomas with 67.3% to 83.7% sensitivity

Insights Imaging · Nov 18, 2026 · Research · Radiology & Imaging

Advancing placenta accreta spectrum diagnosis: is 3-T MRI or combined MRI with ultrasound the answer?

Insights Imaging · Nov 18, 2026 · Research · Radiology & Imaging

Classic imaging signs aid diagnosis in head and neck radiology

Insights Imaging · Nov 15, 2026 · Research · Radiology & Imaging

Automated carotid plaque quantification helps identify obstructive coronary artery disease