ResearchInsights into Imaging · Jan 26, 2026
Prospective validation of an AI software for detecting clinically significant prostate cancer on biparametric MRI
Sunoqrot MRS, Segre R, Nketiah GA, Davik P, Sjøbakk TAE, Langørgen S, et al
Abstract
Objectives
To evaluate the feasibility and safety (primary endpoints), and performance (secondary endpoint) of a new artificial intelligence (AI) software for detecting clinically significant prostate cancer (csPCa) on biparametric MRI (bpMRI) compared to an expert radiologist. Materials and methods
In this prospective study at St. Olavs Hospital, Norway (December 2023-October 2024), 89 consecutive biopsy-naïve men underwent bpMRI for suspected PCa. Scans were interpreted by a radiologist using PI-RADS v2.1 and a radiomics-based AI software. Biopsies were obtained from all radiologist- and/or AI-identified lesions. csPCa was defined as ISUP ≥ 2. Feasibility was defined by a Results
Among 89 patients eligible for primary endpoints evaluation, the software demonstrated feasibility (7% failure rate) and safety (no SADEs). Among 76 eligible for secondary endpoint evaluation (median age 68 years [IQR: 63-73]), csPCa was found in 51% (39/76). Patient-level, software achieved an area under the ROC curve [95% CI] of 0.90 [0.83, 0.96] versus 0.86 [0.76, 0.93] (p = 0.25). At a retrospectively optimized threshold matching the radiologist's patient-level sensitivity at PI-RADS 3 (0.92), software achieved specificity of 0.68 [0.57, 0.78] versus 0.57 [0.46, 0.68] (p = 0.29). Lesion-level, software achieved higher average precision (0.61 [0.52, 0.71] vs. 0.56 [0.46, 0.67]) and lower average false-positive per patient (0.33 [0.22, 0.43] vs. 0.41 [0.30, 0.52]) at the optimized threshold. Conclusion
The software was feasible and safe, and diagnostic performance showed potential to reduce unnecessary biopsies. Critical relevance statement
This clinically validated artificial intelligence software enables feasible and safe detection of clinically significant prostate cancer on biparametric MRI, with demonstrated potential to reduce unnecessary biopsies and improve diagnostic accuracy, indicating potential for integration into clinical prostate cancer care. Key points
A fully automated radiomics software for clinically significant prostate cancer detection on biparametric MRI was prospectively clinically validated. The software demonstrated feasibility and safety, with potential to reduce unnecessary biopsies and improve diagnostic accuracy. The investigated radiomics software has the potential for integration into clinical prostate cancer care.
DOI: 10.1186/s13244-025-02199-9 · PubMed: 41586851
Spotted a mistake? Suggest a correction
More in Radiology & Imaging
ResearchInsights Imaging · Nov 20, 2026 · Radiology & Imaging
Enhanced prognosis stratification following surgery for pancreatic ductal adenocarcinoma to optimize clinical decisions by reflecting tumor heterogeneity and microenvironment
ResearchInsights Imaging · Nov 19, 2026 · Radiology & Imaging
Diagnostic performance of dual-phase photon-counting detector for localization of parathyroid adenomas after inconclusive first-line imaging
ResearchInsights Imaging · Nov 18, 2026 · Radiology & Imaging
ResearchInsights Imaging · Nov 18, 2026 · Radiology & Imaging
Classic signs in head and neck imaging-a quick guide for the radiology resident
ResearchInsights Imaging · Nov 15, 2026 · Radiology & Imaging
Carotid CTA with automated plaque quantification for risk stratification of coexisting obstructive coronary artery disease