Real-World Performance of the Afirma Genomic Sequencing Classifier (GSC)-A Meta-analysis
Key result: Real-world performance of Afirma genomic sequencing classifier was higher than original validation study metrics
Meta-analysisBenefit found
How does real-world Afirma Genomic Sequencing Classifier performance compare to the validation study?
Who
Patients with indeterminate thyroid nodule cytology
Compared
Real-world studies: sensitivity 97%, observed specificity 88%, observed positive predictive value 65%, negative predictive value 99%
Validation study: sensitivity 91%, specificity 68%, positive predictive value 47%, negative predictive value 96%
Main outcome
Diagnostic performance metrics
Bottom line: Real-world performance shows significantly better observed specificity and positive predictive value than the original validation study.
J Clin Endocrinol Metab · 2022-12-06 · Save PNG
Key points
- Researchers performed a meta-analysis of 13 independent real-world studies to compare Afirma genomic sequencing classifier performance against its 2018 validation study.
- Real-world observed specificity was 88% and observed positive predictive value was 65%, which were significantly higher than the validation study metrics.
- The real-world benign call rate was 67%, compared to 54% in the validation study.
Discussion prompts
- Would these patients look like yours? Who was left out?
- Is the main outcome one that matters to patients?
- Is the difference large enough to change what you do?
Cite: Nasr CE, Andrioli M, Endo M, et al. Real-World Performance of the Afirma Genomic Sequencing Classifier (GSC)-A Meta-analysis. J Clin Endocrinol Metab. 2023;108(6):1526-1532. doi:10.1210/clinem/dgac688