ResearchEJNMMI Research · Dec 10, 2025

Renal dynamic scintigraphy using 3D-ring CZT cameras: comparison with NaI(Tl)

Metrard G, Rouzic GL, Chapelle B, Darsin-Bettinger D, Dekeyser M, Bailly M
Radiology & ImagingNephrologyAdult and pediatric
Read the full paper free at PubMed Central ↗

Abstract

Background Isotopic renography is pivotal for assessing upper urinary tract obstruction and provides renal functional parameters. New 3D-CZT cameras offer enhanced sensitivity and innovative full SPECT protocols that need to be validated on phantoms. Current renal phantoms are often complex and suitable for 2D imaging. Using a new printed 3D renal phantom, we compared planar and full SPECT dynamic imaging. Methods A 3D-printed renal phantom mimicking pediatric/adult kidneys was designed. Each kidney subunit, connected to a bladder-bag, was programmably infused with 7.5 MBq of [99mTc]-Pertechnetate to simulate normal nephrograms. An asymmetry of renal activity was simulated using different activity ratios. Each acquisition was conducted on both conventional NaI(Tl) and 3D-CZT cameras. Results The 3D printed phantom enabled the acquisition of consistent and reproducible nephrograms. Tmax, T1/2, and A20 - min/max ratios showed no statistically significant differences between planar and SPECT. Simulated asymmetric renal function demonstrated a strong linear correlation between activity and AUC with r = 0.9455 and 0.9471 (p < 0.0001), respectively for conventional and 3D cameras. Conclusion This innovative 3D phantom allowed the acquisition of reproducible nephrograms with parameters comparable to those of clinical examinations. At low activity, 3D-CZT camera acquisitions provided equivalent values to conventional cameras, supporting their use for qualitative imaging and follow-up.

DOI: 10.1186/s13550-025-01351-2 · PubMed: 41369969

Cite this article

Metrard G, Rouzic GL, Chapelle B, Darsin-Bettinger D, Dekeyser M, Bailly M. Renal dynamic scintigraphy using 3D-ring CZT cameras: comparison with NaI(Tl). EJNMMI Res. 2025. doi:10.1186/s13550-025-01351-2

RIS (EndNote, Zotero, Mendeley)BibTeX

Spotted a mistake? Suggest a correction

Top studies every Monday

The week’s key trials and guidelines in your specialties, plus any topics you follow. By email, free.

Get the weekly digest

More in Radiology & Imaging

Research · Insights Imaging · Nov 20, 2026 · Radiology & Imaging

Integrated system predicts disease-free survival in pancreatic cancer patients better

Meta-analysis · Eur Radiol Exp · Nov 20, 2026 · Radiology & Imaging, Oncology

MRI radiomics of pancreatic tumours: a systematic review of methodological quality and imaging reproducibility

Research · Insights Imaging · Nov 19, 2026 · Radiology & Imaging

Dual-phase photon-counting detector localized parathyroid adenomas with 67.3% to 83.7% sensitivity

Research · EJNMMI Res · Nov 19, 2026 · Radiology & Imaging

Small language model pipeline structures clinical information for dopamine transporter SPECT imaging reports

Research · Insights Imaging · Nov 18, 2026 · Radiology & Imaging

MRI detected placenta accreta spectrum with 100 percent sensitivity