TAVR and surgery showed similar valve durability at seven years
Seven-Year Valve Durability With Transcatheter or Surgical Aortic Valve Replacement: An Ad Hoc Analysis of the PARTNER 3 Randomized Clinical Trial
Why it matters: Limited data exist on long-term valve durability with TAVR or surgery
- 1000 low-risk patients with severe aortic stenosis were randomized to TAVR or surgery.
- 7.3% of TAVR patients and 7.6% of surgery patients had valve dysfunction due to structural deterioration.
- Thrombosis-related valve dysfunction occurred more frequently with TAVR, at 5.2% vs 0.9% with surgery.
Limitation the authors note: This was an ad hoc analysis of a randomized clinical trial
Headline and key points written with AI from the paper’s abstract, with every number checked against it. Read the paper before relying on them.
Topics
Cite this article
Ternacle J, Hahn RT, Silva I, Alperi A, Giuliani C, Zanuttini A. Seven-Year Valve Durability With Transcatheter or Surgical Aortic Valve Replacement: An Ad Hoc Analysis of the PARTNER 3 Randomized Clinical Trial. JAMA Cardiol. 2026. doi:10.1001/jamacardio.2026.2299