Sugammadex interference with serum progesterone measurement: an in-vitro study
Study type: Observational study: it can show a link, but not prove that one thing causes another.
Abstract
Introduction
Sugammadex may interfere with serum progesterone immunoassays, potentially rendering measured concentrations and resulting recommendations unreliable. Methods
We investigated the effect of a range of sugammadex concentrations on serum progesterone concentrations in samples from pre-menopausal female volunteers. Samples were spiked with sugammadex concentrations (15-120 μM), corresponding to clinical doses between 2 mg.kg-1 and 19 mg.kg-1. Total progesterone concentrations in each sample were determined using the routine electrochemiluminescence immunoassay method. A subgroup of 15 samples underwent further testing using liquid chromatography-tandem mass spectroscopy, comparing baseline with sugammadex 120 μM. Results
Thirty-five participants provided blood samples. Increasing concentrations of sugammadex were associated with lower measured serum progesterone, with an approximate 10% reduction with each doubling of sugammadex concentration (F(2,30) = 4.86, p = 0.015). Overall, progesterone concentrations decreased by a mean (SD) of 34.6% (12.4%) between control samples and those spiked with sugammadex 120 μM (p = 0.015). In contrast, liquid chromatography-tandem mass spectroscopy showed no statistically significant biochemical change in progesterone concentrations between baseline and sugammadex 120 μM. Discussion
Sugammadex produced a dose-dependent reduction in the progesterone concentrations when measured by immunoassay techniques, but not by liquid chromatography-tandem mass spectroscopy. This suggests that an immunoassay interference occurs across clinically relevant doses. Progesterone results from routine laboratory testing should be interpreted with caution in the presence of sugammadex.
Topics
Cite this article
Devoy T, Larkin T, Allan M, Mantik K, McShane M, Smith NA. Sugammadex interference with serum progesterone measurement: an in-vitro study. Anaesthesia. 2026. doi:10.1111/anae.70292