ResearchClinical Chemistry · Jul 5, 2026

Interlaboratory Comparison of a Glucagon and Oxyntomodulin Immuno-LC-MS/MS Assay: Implications for Diabetes Research

Moradian A, Becker JO, Nierves LA, Lin TT, Sechi S, MacCoss MJ, et al
Pathology & Laboratory MedicineEndocrinology & DiabetesAdult
Read the full paper free at the publisher ↗

Abstract

Background The quantification of plasma glucagon and oxyntomodulin is important in the assessment of α-cell function, which is impaired in patients with diabetes. We aimed to transfer between laboratories a novel assay that uses liquid chromatography-tandem mass spectrometry (LC-MS/MS) for the sensitive and specific measurement of these highly homologous hormones. Methods Simultaneous measurement of glucagon and oxyntomodulin used immunoaffinity enrichment and LC-MS/MS. Immunoenrichment used monoclonal antibodies that are available at-cost to researchers (deposited at the Developmental Studies Hybridoma Bank). Pure synthetic glucagon, characterized for purity and concentration, was used as a calibrant and is available to others. A detailed standard operating procedure was shared between 3 laboratories and the performance of the method was evaluated with samples collected from patients with and without diabetes. Method comparison was made with 2 commercially available, FDA-registered glucagon immunoassays. Results The method was linear over the normal range (1-20 pM). When measured in duplicate, the median interlaboratory imprecision (%CV) of the measurement of 40 samples was 6.3% (IQR 4.5%) and 14.4% (IQR 12.4%) for glucagon and oxyntomodulin, respectively. Method comparison with commercially available immunoassays demonstrated good (Mercodia, R = 0.92) or fair (Ansh, R = 0.70) agreement. Multivariable linear regression using LC-MS/MS glucagon and oxyntomodulin concentrations to predict immunoassay results indicated significant cross-reactivity of each immunoassay with oxyntomodulin. Conclusion We have validated a sensitive and specific assay for glucagon and oxyntomodulin that can be deployed in high-complexity clinical laboratories for research or the care of patients. Commercially available glucagon immunoassays have significant interference from molecules other than glucagon.

Subject terms

Glucagon; Humans; Oxyntomodulin; Tandem Mass Spectrometry; Liquid Chromatography-Mass Spectrometry; Immunoassay; Diabetes Mellitus; Chromatography, Liquid; Reproducibility of Results

DOI: 10.1093/clinchem/hvag078 · PubMed: 42400596

Headline, design, result, key points, caveat and citation as plain text.

Cite this article

Moradian A, Becker JO, Nierves LA, et al. Interlaboratory Comparison of a Glucagon and Oxyntomodulin Immuno-LC-MS/MS Assay: Implications for Diabetes Research. Clin Chem. 2026;72(11):1161-1170. doi:10.1093/clinchem/hvag078

RIS (EndNote, Zotero, Mendeley)BibTeX

Spotted a mistake? Suggest a correction

Top Pathology & Laboratory Medicine studies every Monday

The week’s key trials and guidelines, free to read, in one short email. No account; unsubscribe in one click.

Choose specialties and topics

Similar articles

Research · Clin Chem · Jan 7, 2026 · Pathology & Laboratory Medicine

Development of a Novel Liquid Chromatography Coupled to Multiple Reaction Monitoring (LC-MRM) Assay for the Quantification of Neurofilament Light Chain in Cerebrospinal Fluid and Comparison with Ultra-Sensitive Immunoassay: A Step toward Standardization

Research · Clin Chem · Apr 18, 2026 · Pathology & Laboratory Medicine

Transferability and Between Laboratory Imprecision of an LC-MS/MS-Based Reference Measurement Procedure for Apolipoproteins

Research · Clin Chem · Jun 17, 2026 · Pathology & Laboratory Medicine

Correlation and Commutability Study between an LC-MS-Based Reference Measurement Procedure and Immunoassays Measuring Apolipoprotein A-I and Apolipoprotein B

Research · J Clin Endocrinol Metab · Feb 19, 2026 · Endocrinology & Diabetes

Evaluation of the Siemens Healthineers Atellica TSTII testosterone assay in the CDC HoSt-TT program

Research · Clin Chem · May 27, 2026 · Pathology & Laboratory Medicine

Development and Analytical Validation of a Multiplex LC-MRM-MS-Based Reference Measurement Procedure for Apolipoprotein A-I and Btotal Quantification in Serum