Interlaboratory Comparison of a Glucagon and Oxyntomodulin Immuno-LC-MS/MS Assay: Implications for Diabetes Research
Abstract
Background
The quantification of plasma glucagon and oxyntomodulin is important in the assessment of α-cell function, which is impaired in patients with diabetes. We aimed to transfer between laboratories a novel assay that uses liquid chromatography-tandem mass spectrometry (LC-MS/MS) for the sensitive and specific measurement of these highly homologous hormones. Methods
Simultaneous measurement of glucagon and oxyntomodulin used immunoaffinity enrichment and LC-MS/MS. Immunoenrichment used monoclonal antibodies that are available at-cost to researchers (deposited at the Developmental Studies Hybridoma Bank). Pure synthetic glucagon, characterized for purity and concentration, was used as a calibrant and is available to others. A detailed standard operating procedure was shared between 3 laboratories and the performance of the method was evaluated with samples collected from patients with and without diabetes. Method comparison was made with 2 commercially available, FDA-registered glucagon immunoassays. Results
The method was linear over the normal range (1-20 pM). When measured in duplicate, the median interlaboratory imprecision (%CV) of the measurement of 40 samples was 6.3% (IQR 4.5%) and 14.4% (IQR 12.4%) for glucagon and oxyntomodulin, respectively. Method comparison with commercially available immunoassays demonstrated good (Mercodia, R = 0.92) or fair (Ansh, R = 0.70) agreement. Multivariable linear regression using LC-MS/MS glucagon and oxyntomodulin concentrations to predict immunoassay results indicated significant cross-reactivity of each immunoassay with oxyntomodulin. Conclusion
We have validated a sensitive and specific assay for glucagon and oxyntomodulin that can be deployed in high-complexity clinical laboratories for research or the care of patients. Commercially available glucagon immunoassays have significant interference from molecules other than glucagon.
Subject terms
Headline, design, result, key points, caveat and citation as plain text.
Cite this article
Moradian A, Becker JO, Nierves LA, et al. Interlaboratory Comparison of a Glucagon and Oxyntomodulin Immuno-LC-MS/MS Assay: Implications for Diabetes Research. Clin Chem. 2026;72(11):1161-1170. doi:10.1093/clinchem/hvag078
Spotted a mistake? Suggest a correction