Conventional radiation protection provided better chest and abdomen shielding than the Zero Gravity system
The Effect of a Suspended Radiation Protection System on Occupational Radiation Doses During Infrarenal EVAR Procedures: A Randomised Controlled Study
Prospective, randomized, two arm trialMixed results
Does Zero Gravity radiation protection compare to conventional protection during infrarenal EVAR?
Who
71
Consecutive elective infrarenal endovascular aneurysm repair procedures · Single centre
Compared
Zero Gravity operator and conventional assistant: axilla 1.5 μSv and eyes 6.3 μSv
Conventional protection for operator and assistant: chest 0.1 μSv, abdomen 0.6 μSv, lower leg 2.2 μSv
Main outcome
Deep dose equivalent and personal dose equivalent values
Bottom line: Zero Gravity improved axilla and eye shielding but conventional protection was superior for chest, abdomen, and lower leg.
Eur J Vasc Endovasc Surg · 2023-08-22 · NCT04078165
Limitation the authors note: The study was a single centre, prospective, randomised, two arm trial.
Headline and visual abstract written with AI from the paper’s abstract, with every number checked against it. Read the paper before relying on them.