ResearchJAMA · Jan 10, 2023

Use of Recalled Devices in New Device Authorizations Under the US Food and Drug Administration's 510(k) Pathway and Risk of Subsequent Recalls

Kadakia KT, Dhruva SS, Caraballo C, Ross JS, Krumholz HM
General MedicineAdult

Study type: Observational study: it can show a link, but not prove that one thing causes another.

Read the full paper free at PubMed Central ↗

Abstract

Importance In the US, nearly all medical devices progress to market under the 510(k) pathway, which uses previously authorized devices (predicates) to support new authorizations. Current regulations permit manufacturers to use devices subject to a Class I recall-the FDA's most serious designation indicating a high probability of adverse health consequences or death-as predicates for new devices. The consequences for patient safety are not known. Objective To determine the risk of a future Class I recall associated with using a recalled device as a predicate device in the 510(k) pathway. Design and setting In this cross-sectional study, all 510(k) devices subject to Class I recalls from January 2017 through December 2021 (index devices) were identified from the FDA's annual recall listings. Information about predicate devices was extracted from the Devices@FDA database. Devices authorized using index devices as predicates (descendants) were identified using a regulatory intelligence platform. A matched cohort of predicates was constructed to assess the future recall risk from using a predicate device with a Class I recall. Main outcomes and measures Devices were characterized by their regulatory history and recall history. Risk ratios (RRs) were calculated to compare the risk of future Class I recalls between devices descended from predicates with matched controls. Results Of 156 index devices subject to Class I recall from 2017 through 2021, 44 (28.2%) had prior Class I recalls. Predicates were identified for 127 index devices, with 56 (44.1%) using predicates with a Class I recall. One hundred four index devices were also used as predicates to support the authorization of 265 descendant devices, with 50 index devices (48.1%) authorizing a descendant with a Class I recall. Compared with matched controls, devices authorized using predicates with Class I recalls had a higher risk of subsequent Class I recall (6.40 [95% CI, 3.59-11.40]; P<.001). Conclusions and relevance Many 510(k) devices subjected to Class I recalls in the US use predicates with a known history of Class I recalls. These devices have substantially higher risk of a subsequent Class I recall. Safeguards for the 510(k) pathway are needed to prevent problematic predicate selection and ensure patient safety.

Subject terms

Humans; Cross-Sectional Studies; Databases, Factual; Device Approval; Medical Device Recalls; United States; United States Food and Drug Administration

DOI: 10.1001/jama.2022.23279 · PubMed: 36625810

Cite this article

Kadakia KT, Dhruva SS, Caraballo C, Ross JS, Krumholz HM. Use of Recalled Devices in New Device Authorizations Under the US Food and Drug Administration's 510(k) Pathway and Risk of Subsequent Recalls. JAMA. 2023. doi:10.1001/jama.2022.23279

RIS (EndNote, Zotero, Mendeley)BibTeX

Spotted a mistake? Suggest a correction

Top General Medicine studies every Monday

The week’s key trials and guidelines, free to read, in one short email. No account; unsubscribe in one click.

Choose specialties and topics

Similar articles

Research · JAMA · Jan 10, 2023 · Digital Health & AI

Association Between Regulatory Submission Characteristics and Recalls of Medical Devices Receiving 510(k) Clearance

Research · JAMA Health Forum · Aug 7, 2026

Premarket Clinical Evidence Strength and Recalls of High-Risk Therapeutic Medical Devices

Research · JAMA Netw Open · Jun 1, 2026 · Digital Health & AI

Clinical Evidence and FDA Recalls of Artificial Intelligence-Enabled Medical Devices

Research · BMJ · Oct 14, 2026

Most adverse event reports to FDA were incomplete or delayed

Research · BMJ · Mar 12, 2025

Late adverse event reporting from medical device manufacturers to the US Food and Drug Administration: cross sectional study