ResearchEuropean Radiology Experimental · Sep 30, 2026

Accuracy of iodinated contrast and saline delivery using a piston-driven and a peristaltic power injector in CT-an in vitro study

Alyani Nezhad Z, Salyapongse AM, Goetsch CJ, Bladorn RM, Bartels CM, Schluter KL, et al
Radiology & ImagingBasic science

Study type: Laboratory or animal research: results may not apply to people yet.

Read the full paper free at PubMed Central ↗

Abstract

Objective To evaluate the accuracy of iodinated contrast media (ICM) and saline flush volumes delivered by a piston- and a peristaltic-based injector in an in vitro study. Materials and methods Seven clinical injection protocols, ranging from pediatric to bariatric indications and incorporating varying flow rates, ICM volumes, saline volumes, and dilution ratios, were tested. Identical prescriptions were programmed into a piston-based injector and a peristaltic-based injector. Injections were performed into vials. Calibration samples (5-20 mgI/mL) were scanned on a multi-slice CT scanner to generate a calibration curve relating CT number to iodine concentration. The saline test volume and combined ICM + saline flush volume were measured for each injection. CT numbers of mixtures were converted to iodine concentration using the calibration curve, and delivered ICM and saline volumes were calculated. Percent errors between expected and measured volumes were determined. A Wilcoxon signed-rank test assessed statistical significance. Results For the peristaltic-based injector, ICM volume error ranged from 4.0% to 10.0% and saline flush error from 0.6% to 9.2%. For the piston-based injector, ICM volume error ranged from 9.9% to 40.0% and saline flush error from 1.5% to 50.2%. Differences between expected and delivered volumes were statistically significant across protocols (p = 0.031/= 0.035), except for peristaltic saline flush protocols (Ped 12 mL (p = 0.094), Adult 170 mL (p = 0.156), and High Flow (p = 0.093)). Conclusions Both injectors delivered excess ICM while saline flush volumes showed variability. These findings underscore the importance of performance verification to ensure accurate and consistent contrast delivery in clinical practice. Key points Question Automated injectors often delivered more contrast than expected. Findings Both systems often underdelivered saline with some instances of overdelivery. Relevance statement Unrecognized variability in contrast and saline delivery across different injection settings may affect enhancement consistency, contrast exposure, and diagnostic reliability. Broader injector performance evaluation across clinically used protocols and clinically simulated setups may improve CT practice.

Subject terms

Contrast Media; Tomography, X-Ray Computed; Humans; Saline Solution; Iodine; Sodium Chloride; Reproducibility of Results; Injections; In Vitro Techniques; Equipment Design

DOI: 10.1186/s41747-026-00785-0 · PubMed: 42814175

Cite this article

Alyani Nezhad Z, Salyapongse AM, Goetsch CJ, Bladorn RM, Bartels CM, Schluter KL. Accuracy of iodinated contrast and saline delivery using a piston-driven and a peristaltic power injector in CT-an in vitro study. Eur Radiol Exp. 2026. doi:10.1186/s41747-026-00785-0

RIS (EndNote, Zotero, Mendeley)BibTeX

Spotted a mistake? Suggest a correction

Top studies every Monday

The week’s key trials and guidelines in your specialties, plus any topics you follow. By email, free.

Get the weekly digest

Similar articles

Meta-analysis · Eur Radiol · Dec 14, 2023 · Radiology & Imaging

Iodine-based contrast media may be linked to increased myasthenia gravis symptoms in few patients

Research · Eur Radiol Exp · Jul 1, 2026 · Radiology & Imaging, Gastroenterology & Hepatology

Optimization of hypovascular liver lesion detectability in dual-energy CT using deep learning image reconstruction: a phantom study for potential iodine dose reduction

Research · Eur Radiol Exp · Jun 25, 2026 · Radiology & Imaging, Gastroenterology & Hepatology

DECT-based stratification of nonocclusive mesenteric ischemia using bowel-wall iodine concentration: a prospective single-center cohort

Research · Eur Radiol Exp · Jun 8, 2026 · Radiology & Imaging, Digital Health & AI

The impact of a novel deep learning reconstruction algorithm on image quality in ultralow-dose CT: a quantitative phantom study

Research · Eur Radiol Exp · Oct 12, 2026 · Radiology & Imaging

Photon-counting CT showed lower iodine detection and quantification thresholds than dual-energy CT systems