Bihormonal artificial pancreas increased time in target glucose range after total pancreatectomy
Bihormonal Artificial Pancreas With Closed-Loop Glucose Control vs Current Diabetes Care After Total Pancreatectomy: A Randomized Clinical Trial
Clinical question: Does bihormonal artificial pancreas improve glucose control after total pancreatectomy?
Why it matters: Glucose control is difficult after total pancreatectomy due to the complete absence of alpha and beta cells, and the efficacy of a bihormonal artificial pancreas in this setting is unknown.
- The study compared a bihormonal artificial pancreas using insulin and glucagon to standard insulin pump or pen therapy in 12 adult outpatients after total pancreatectomy.
- Patients using the bihormonal system spent a median of 78.30% of time in euglycemia, or target glucose range, compared to 57.38% with current diabetes care.
- The bihormonal device was associated with a lower median time in hypoglycemia of 0.00% compared to 1.61% for current diabetes care.
Headline and key points written with AI from the paper’s abstract, with every number checked against it. Read the paper before relying on them.
Was this summary helpful?
Subject terms
Cite this article
van Veldhuisen CL, Latenstein AEJ, Blauw H, Vlaskamp LB, Klaassen M, Lips DJ. Bihormonal Artificial Pancreas With Closed-Loop Glucose Control vs Current Diabetes Care After Total Pancreatectomy: A Randomized Clinical Trial. JAMA Surg. 2022. doi:10.1001/jamasurg.2022.3702
Spotted a mistake? Suggest a correction