Comparing the detectability of pulmonary nodules on two ultra-high resolution CT scanners: a preliminary phantom study

Greffier J, Salvat C, Pastor M, Villani N, Bousson V, Vallot A, et al
Radiology & ImagingPulmonologyAdult
Read the full paper free at PubMed Central ↗
Top Radiology & Imaging studies every Monday

The week’s key trials and guidelines, free to read, in one short email. No account; unsubscribe in one click.

Choose specialties and topics

Abstract

Objectives To objectively and subjectively compare the detectability of pulmonary nodules on an ultra-high-resolution CT (UHR-CT) scanner equipped with energy-integrating detectors and a photon-counting CT (PCCT) in UHR mode. Materials and methods An image quality phantom and an anthropomorphic phantom were scanned on a UHR-CT and a PCCT in UHR mode at 7.5/2.5/0.4 mGy. The parameters typically used for UHR chest CT scans were selected. Noise power spectrum and task-based transfer function were computed to assess noise-magnitude, noise texture (fav), and spatial resolution (f50), respectively. Detectability indices (d') were computed to model the detection of three chest nodules. Five radiologists subjectively evaluated their confidence in detecting these nodules available on the anthropomorphic phantom. Results For PCCT, noise magnitude increased, and fav values were similar as the dose decreased. For UHR-CT, noise-magnitude decreased, and fav values decreased as the dose decreased. For both inserts and CT systems, f50 values decreased as the dose decreased. For both inserts and at all dose levels, f50 values were higher with PCCT than with UHR-CT. d' values increased as the dose increased with PCCT, and the opposite was found with UHR-CT. Confidence in nodule detection was considered clinically sufficient at all dose levels for high-contrast solid nodules at 7.5 and 2.5 mGy for subsolid nodules, and only for PCCT at 7.5 mGy for low-contrast nodules. Conclusion Under UHR conditions, pulmonary nodule detectability strongly depends on the dose level and reconstruction parameters. High-contrast nodule detection can be performed at all dose levels with both CT systems, but reconstruction strategies must be optimized for low-contrast nodule detection. Key points Do the performance of UHR-CT and PCCT scanners improve the detectability of pulmonary nodules? PCCT showed dose-invariant noise texture and higher effective spatial resolution, whereas UHR-CT relied on dose-dependent noise suppression, producing smoother images and reduced spatial frequency. High-contrast nodule detection can be performed at all dose levels with both CT systems, particularly at ultra-low dose levels, but reconstruction strategies must be optimized for low-contrast nodule detection.

More on this topic

Lung cancer

Subject terms

Phantoms, Imaging; Tomography, X-Ray Computed; Humans; Solitary Pulmonary Nodule; Reproducibility of Results; Sensitivity and Specificity; Tomography Scanners, X-Ray Computed; Radiation Dosage; Lung Neoplasms; Equipment Design; Equipment Failure Analysis; Radiographic Image Interpretation, Computer-Assisted

DOI: 10.1007/s00330-026-12582-9 · PubMed: 42056488

Cite this article

Greffier J, Salvat C, Pastor M, et al. Comparing the detectability of pulmonary nodules on two ultra-high resolution CT scanners: a preliminary phantom study. Eur Radiol. 2026. doi:10.1007/s00330-026-12582-9

RIS (EndNote, Zotero, Mendeley)BibTeX

Spotted a mistake? Suggest a correction

Similar articles

Guideline · Eur Radiol · Jul 1, 2025 · Radiology & Imaging, Pulmonology

European society recommends measuring nodule growth to guide lung cancer screening follow-up

Research · Eur Radiol · Apr 17, 2026 · Radiology & Imaging, Pulmonology

Features of low-dose CT-detected lung nodules: individuals who never smoked vs. who smoke(d) in a Chinese general population

Research · Eur Radiol · Feb 19, 2026 · Radiology & Imaging, Cardiology

Coronary artery calcium scoring: expanding the new standard by photon-counting detector CT-Part I: Impact of tube voltage, tube current, slice thickness, and quantum iterative reconstructions

Research · Eur Radiol · Feb 19, 2026 · Radiology & Imaging, Cardiology

Coronary artery calcium scoring: expanding the new standard by photon-counting detector CT Part II: Impact of virtual monoenergetic image reconstructions with adjusted calcium scoring thresholds

Research · Respir Res · Feb 11, 2026 · Pulmonology, Radiology & Imaging

Ultra-high-resolution synchrotron phase-contrast CT enables microstructural pulmonary imaging at clinical dose levels