Abstract

Objectives Part I of this study introduced a new photon-counting detector (PCD-)CT protocol for coronary artery calcium (CAC) scoring using 120 kVp, 75% dose, thin slices, quantum iterative reconstructions (IR) 2, leading to a significant reduction of score variability. The second part evaluated the potential of virtual monoenergetic image (VMI) reconstruction in further reducing score variability with PCD-CT. Materials and methods CAC scoring was performed on PCD-CT with a chest phantom containing nine calcifications using the optimized PCD-CT protocol from Part I. Images were reconstructed at different VMI levels (50-80 keV, 5 keV-steps), with adjusted CAC thresholds to maintain density equivalence to 70 keV. CAC scores, image noise, and calcification detectability were investigated. Results were compared to standard PCD-CT, EID-CT and previously proposed EID-CT protocols. Results Using 65 keV reconstructions, score variability decreased by 9% compared to the optimized PCD-CT protocol from Part I, by 43% vs. the standard PCD-CT, by 78% vs. the standard EID-CT, and by 69% vs. the proposed EID-CT protocol. Image noise remained within targets, eliminating the risk of false-positives. Calcification detectability was comparable to the optimized PCD-CT protocol (7.1 ± 0.6 vs. 7.1 ± 0.8). Calcium volume and mass scores from the keV-optimized PCD protocol were closer to the physical reference compared to scores from the standard PCD protocol. Conclusions Score variability and calcification detectability in PCD-CT-based CAC scoring can be further improved when augmenting an optimized PCD-CT protocol at 65 keV. In addition to reducing the radiation dose, this protocol may enable more consistent CAC quantification and seems to perform even better than the proposed, multivendor EID-CT protocol. Key points Question Coronary calcium scoring lacks reproducibility. Adding virtual monoenergetic imaging with adapted thresholds to a pre-optimized photon-counting CT protocol may further improve score variability. Findings A 120 kVp, 75%-dose, thin-slice photon-counting CT protocol at 65 keV achieved the lowest coronary calcium score variability compared to previous protocols. Clinical relevance Minimizing variability in coronary calcium scoring improves the technical reliability of serial measurements. The additional use of virtual monoenergetic imaging further reduces variability in photon-counting CT, supporting a precise and consistent cardiovascular risk assessment.

Subject terms

Humans; Phantoms, Imaging; Coronary Artery Disease; Radiographic Image Interpretation, Computer-Assisted; Photons; Reproducibility of Results; Tomography, X-Ray Computed; Coronary Angiography; Calcinosis; Sensitivity and Specificity; Algorithms; Calcium

DOI: 10.1007/s00330-026-12356-3 · PubMed: 41714425

Cite this article

Fink N, Koetzier LR, Zsarnoczay E, et al. Coronary artery calcium scoring: expanding the new standard by photon-counting detector CT Part II: Impact of virtual monoenergetic image reconstructions with adjusted calcium scoring thresholds. Eur Radiol. 2026;36(7):5557-5566. doi:10.1007/s00330-026-12356-3

RIS (EndNote, Zotero, Mendeley)BibTeX

Spotted a mistake? Suggest a correction

Top Radiology & Imaging studies every Monday

The week’s key trials and guidelines, free to read, in one short email. No account; unsubscribe in one click.

Choose specialties and topics

Similar articles

Research · Eur Radiol · Feb 19, 2026 · Radiology & Imaging, Cardiology

Coronary artery calcium scoring: expanding the new standard by photon-counting detector CT-Part I: Impact of tube voltage, tube current, slice thickness, and quantum iterative reconstructions

Research · Eur Radiol Exp · Sep 30, 2026 · Radiology & Imaging, Oncology

Optimization of spectral photon-counting CT reconstruction parameters for improved detection of peritoneal metastases compared with energy-integrating detector dual-energy CT: a phantom study

Research · Eur Radiol Exp · May 20, 2026 · Radiology & Imaging, Cardiology

Effect of quantum iterative reconstruction on coronary stent assessment in ultra-high resolution photon-counting detector CT: a phantom study

Research · Eur Radiol · Apr 29, 2026 · Radiology & Imaging, Pulmonology

Comparing the detectability of pulmonary nodules on two ultra-high resolution CT scanners: a preliminary phantom study

Research · Eur Radiol Exp · May 6, 2026 · Radiology & Imaging, Cardiology

Lumen imaging in calcified coronary arteries using high-resolution Gd-enhanced color K-edge imaging with spectral photon-counting CT: a phantom study