Two-person mask ventilation was associated with lower mask leak in preterm infants at birth
One-person versus two-person mask ventilation in preterm infants at birth: a pilot randomised controlled trial
Pilot randomized controlled trialEarly signal
Does two-person mask ventilation improve respiratory mechanics compared to one-person mask ventilation?
Who
25
Preterm infants less than 30 weeks' gestation
Compared
One-person mask method: 26.4±18.5% mask leak
Two-person mask method: 17.6±9.3% mask leak
Main outcome
Percentage mask leak
Bottom line: Two-person mask ventilation resulted in lower percentage mask leak in preterm infants compared to one-person ventilation.
BMJ Paediatr Open · 2023-02-06 · ACTRN12614000245695
Why it matters: Mask leak and airway obstruction during resuscitation can lead to suboptimal ventilation in newborn infants.
Headline and visual abstract written with AI from the paper’s abstract, with every number checked against it. Read the paper before relying on them.
Was this summary helpful?
Abstract
Background
Mask leak and airway obstruction are common with mask ventilation in newborn infants, leading to suboptimal ventilation. We aimed to perform a pilot study measuring respiratory mechanics during one-person and two-person mask ventilation in preterm infants at birth. Methods
Infants less than 30 weeks' gestation were eligible for the study. In the two-person method, one person holds the mask in place and the other provides positive pressure ventilation compared with the standard one-person mask hold. A respiratory function monitor was used in line with a T-piece resuscitator to measure mask leak and airway obstruction. Deferred consent was obtained. Results
Twenty-five infants were recruited. The mean (SD) birth weight was 920.4 g (188.3), and mean (SD) gestational age was 27.3 weeks (3.0). Percentage mask leak was higher in the one-person mask method (26.4±18.5) compared with the two-person mask method (17.6±9.3) (p=0.018). The mean (SD) expired tidal volume (VTe, mL) in breaths with leak was 3.9 (1.57) in the one-person method compared with 3.05 (1.0) the two-person method (p=0.31). A significantly lower mean (SD) end-tidal carbon dioxide (EtCO2, mm Hg) was measured at 25.3 (9.9) in breaths with mask leak, compared with 30.8 (12.1) in breaths without leak. The breaths with airway obstruction had lower mean EtCO2 (25.9 vs 30.8, p=0.003) and lower mean VTe (1.71 vs 6.95, p<0.001). Conclusion
Mask leak and airway obstruction are common in resuscitation of preterm infants at birth. The use of the two-person mask technique is effective and it could be a useful option if mask ventilation with the one-person method is not effective. Trial registration number
ACTRN12614000245695.