Electrode placement affects spinal reflex thresholds during transcutaneous spinal stimulation in healthy adults
Optimizing transcutaneous spinal stimulation: excitability of evoked spinal reflexes is dependent on electrode montage
Why it matters: It is unknown how different electrode positions for transcutaneous spinal stimulation influence the activation of spinal circuits.
- Fifteen adults with intact neurological systems were studied in a randomized crossover trial comparing six different electrode montages for transcutaneous spinal stimulation.
- The DV-U and DV-I montages were associated with lower reflex thresholds for dominant (D: 46.7[33.9, 59.4], ND: 45.4[32.5, 58.2]) and nondominant (D: 48.1[35.3, 60.8], ND: 45.4[32.5, 58.2]) soleus muscles.
- No differences were found in response amplitude at 120% of reflex threshold or in the tolerability of the stimulation among the different electrode montages.
Headline and key points written with AI from the paper’s abstract, with every number checked against it. Read the paper before relying on them.
Topics
Cite this article
Thatcher KL, Nielsen KE, Sandler EB, Daliet OJ, Iddings JA, Field-Fote EC. Optimizing transcutaneous spinal stimulation: excitability of evoked spinal reflexes is dependent on electrode montage. J Neuroeng Rehabil. 2025. doi:10.1186/s12984-024-01524-5
Spotted a mistake? Suggest a correction
The week’s key trials and guidelines in your specialties, by email. Free.
Get the weekly digest