ResearchEuropean Radiology Experimental · Apr 2, 2026

Phantom-derived method for improving accurate material decomposition in photon-counting detector CT

Springer S, Martens B, Flohr T, Jeukens C
Radiology & ImagingAdult
Read the full paper free at PubMed Central ↗

Abstract

Objective Photon-counting detector CT (PCD-CT) enables spectral imaging with material separation. Accurate iodine and iron quantification remains challenging due to inevitable low- and high-energy base material CT number mismatches and dual-energy ratio (DER) variability. This study develops and validates a correction method addressing these issues to improve iodine and iron quantification in PCD-CT. Materials and methods A spectral CT abdomen phantom containing rods with known iodine (0.5-15.0 mg/mL) and iron (2.0-25.0 mg/mL) concentrations in water- and liver-equivalent material was scanned on a clinical PCD-CT under varying tube voltages, dose levels, and with/without a fat ring. High- and low-energy CT numbers of base materials and DER values were inputs for the correction method. Material concentrations calculated with and without correction were validated against known phantom values. Results The correction method significantly reduced quantification errors. Iodine errors fell below 5% for concentrations ≥ 2 mg/mL and iron errors below 15% for concentrations ≥ 5 mg/mL. Without correction, errors reached up to 83% (iodine) and 85% (iron) at low concentrations, reduced to 23% and 47%, respectively, after correction. Conclusion The proposed correction method improves accuracy in spectral material decomposition for PCD-CT, supporting its potential for better clinical assessment of lesion contrast enhancement, therapy response and hepatic burden evaluation. Relevance statement This technical note introduces a phantom-based correction method for photon-counting detector CT that improves iodine and iron quantification by addressing base material Hounsfield Unit (HU) mismatches and dual-energy ratio variability. The method reduces quantification errors and offers a practical calibration procedure, supporting the potential for clinically reliable iodine and iron quantification. Key points PCD-CT correction method reduces concentration errors across varying scan protocols and configurations. Implementation guide supports adaptation to other scanners. Accurate iodine and iron quantification supports diagnosis and treatment assessment.

Topics

Phantoms, Imaging; Tomography, X-Ray Computed; Photons; Iron; Iodine; Humans

DOI: 10.1186/s41747-026-00705-2 · PubMed: 41925986

Cite this article

Springer S, Martens B, Flohr T, Jeukens C. Phantom-derived method for improving accurate material decomposition in photon-counting detector CT. Eur Radiol Exp. 2026. doi:10.1186/s41747-026-00705-2

RIS (EndNote, Zotero, Mendeley)BibTeX

Spotted a mistake? Suggest a correction

Similar articles

Eur Radiol Exp · Oct 12, 2026 · Research · Radiology & Imaging

Photon-counting CT showed lower iodine detection and quantification thresholds than dual-energy CT systems

Eur Radiol Exp · Jul 1, 2026 · Research · Radiology & Imaging, Gastroenterology & Hepatology

Optimization of hypovascular liver lesion detectability in dual-energy CT using deep learning image reconstruction: a phantom study for potential iodine dose reduction

Eur Radiol Exp · Jun 8, 2026 · Research · Radiology & Imaging, Digital Health & AI

The impact of a novel deep learning reconstruction algorithm on image quality in ultralow-dose CT: a quantitative phantom study

Eur Radiol Exp · Sep 30, 2026 · Research · Radiology & Imaging, Oncology

Optimization of spectral photon-counting CT reconstruction parameters for improved detection of peritoneal metastases compared with energy-integrating detector dual-energy CT: a phantom study

Eur Radiol Exp · Nov 13, 2026 · Research · Radiology & Imaging

Photon-counting CT was linked to better image quality for tantalum clips in proton therapy