Abstract

Background Drug-coated balloons (DCB) are frequently used during percutaneous coronary intervention. Despite this there are limited data from studies comparing different types of DCB, particularly for a paclitaxel-coated balloon (PCB) using urea as the excipient. This study aimed to assess the clinical efficacy of the PCB with urea compared with other frequently used DCB, attempting to emulate a target trial using observational data. Methods SCAAR (Swedish Coronary Angiography and Angioplasty Registry) is a nationwide registry including data on all patients in Sweden undergoing coronary angiography. For this study, all procedures with DCB in Sweden between August 2021 and May 2024 were included. Propensity score weighted multivariable Cox regression analysis and Kaplan-Meier estimates were used to assess outcomes through 2 years. Results A total of 1797 (27.7%) procedures with the PCB with urea and 4691 (72.3%) procedures with Other DCB were eligible for this analysis. DCB treatment was performed in a variety of patients and lesions including 32.4% diabetes, 64.2% acute coronary syndrome, 20.5% bifurcations and 33.0% in-stent restenosis. After 2 years, the event rates for the PCB with urea and Other DCB were 16.4% versus 18.4% for major adverse cardiovascular events (hazard ratio [HR] 1.01 [95% CI 0.84-1.22]), 7.2% versus 8.2% for all-cause mortality (HR 1.05 [95% CI 0.80-1.39]), 6.1% versus 8.0% for myocardial infarction (HR 0.81 [95% CI 0.59-1.11]), 14.6% versus 14.9% for new revascularization with percutaneous coronary intervention (HR 1.02 [95% CI 0.83-1.25]), 0.8% versus 1.3% for target lesion definite thrombosis (HR 0.93 [95% CI 0.41-2.11]), 7.4% versus 8.7% for target lesion revascularization (HR 0.95 [95% CI 0.70-1.29]), and 10.5% versus 11.3% for target vessel revascularization (HR 0.99 [95% CI 0.76-1.29]). Conclusions In this analysis evaluating 2-year clinical outcomes after DCB, in a broad patient population, the PCB with urea had comparable outcomes with other contemporary DCB. This data further support the safety and efficacy of the PCB with urea.

Abstract reproduced under the article’s CC BY licence.

Subject terms

Humans; Paclitaxel; Urea; Cardiovascular Agents; Coated Materials, Biocompatible; Coronary Angiography; Treatment Outcome; Registries; Equipment Design; Time Factors; Aged; Middle Aged

DOI: 10.1161/jaha.125.045621 · PubMed: 42216271

Headline, design, result, key points, caveat and citation as plain text.

Cite this article

von Koch S, Frank E, Gabrysch K, et al. Real-World Usage of a Paclitaxel-Coated Balloon With Urea Compared With Other Contemporary Drug-Coated Balloons: A 2-Year Analysis From SCAAR in Over 6000 Patients. J Am Heart Assoc. 2026;15(11):e045621. doi:10.1161/jaha.125.045621

RIS (EndNote, Zotero, Mendeley)BibTeX

Spotted a mistake? Suggest a correction

Top Cardiology studies every Monday

The week’s key trials and guidelines, free to read, in one short email. No account; unsubscribe in one click.

Choose specialties and topics

Similar articles

Randomized trial n = 70 · J Am Heart Assoc · Sep 29, 2026 · Cardiology

Sirolimus drug-coated balloons improved endothelial function more than paclitaxel balloons in peripheral artery disease

Systematic review n = 259,137 · JACC · Mar 25, 2024 · Cardiology

Paclitaxel-coated devices not linked to higher death risk in large French cohort

Meta-analysis n = 1,682 · Eur J Vasc Endovasc Surg · Aug 20, 2021 · Cardiothoracic & Vascular Surgery, Cardiology

Paclitaxel coated balloons reduced loss of access patency in dialysis patients

Meta-analysis n = 2,930 · Eur J Vasc Endovasc Surg · Apr 1, 2026 · Cardiothoracic & Vascular Surgery, Cardiology

Paclitaxel coated balloons reduce the risk of vascular access narrowing recurrence compared to plain balloons

Systematic review · Eur J Vasc Endovasc Surg · May 18, 2026 · Cardiothoracic & Vascular Surgery, Cardiology

Paclitaxel coated devices show limited benefit and possible long term harm for leg artery disease