Minimally invasive pulse wave analysis showed high error rates compared to standard cardiac output measurements
Agreement of minimally invasive pulse wave analysis with pulmonary artery and transpulmonary thermodilution cardiac output measurements in perioperative and intensive care medicine: a systematic review and meta-analysis
Clinical question: How accurate is minimally invasive pulse wave analysis compared to thermodilution for cardiac output?
Why it matters: Clinicians need reliable ways to monitor cardiac output, which is the volume of blood the heart pumps per minute, in high-risk surgical and critically ill patients.
- Researchers analyzed 92 studies involving 3111 adult surgical or critically ill patients to compare pulse wave analysis with standard thermodilution methods.
- The pooled percentage error was 44.0% for cardiac output and 49.1% for cardiac index, exceeding the 30% threshold for clinical acceptability.
- The percentage error varied substantially depending on the specific device used and the patient population studied.
Headline and key points written with AI from the paper’s abstract, with every number checked against it. Read the paper before relying on them.
Was this summary helpful?
Subject terms
Headline, design, result, key points, caveat and citation as plain text.
Cite this article
Flick M, Müller DX, Bergholz A, et al. Agreement of minimally invasive pulse wave analysis with pulmonary artery and transpulmonary thermodilution cardiac output measurements in perioperative and intensive care medicine: a systematic review and meta-analysis. Br J Anaesth. 2026;137(1):39-54. doi:10.1016/j.bja.2026.04.045