Most cardiac output monitors show poor agreement with reference methods in septic shock patients
Cardiac output monitors in septic shock: do they deliver what matters? A systematic review and meta-analysis
- Researchers analyzed 26 studies involving 1,323 adult intensive care unit patients to compare different cardiac output monitors against reference methods like pulmonary artery catheters.
- The pooled percentage error was 49%, though calibrated pulse contour analysis showed better agreement with a percentage error of 25%.
- Only 3 of 15 datasets reporting trending ability achieved a concordance of 90% or higher.
Limitation the authors note: The authors note that usual validation frameworks overlook critical metrics like precision, time response, and trending ability that are necessary for real-time decision-making.