ResearchCritical Care · Nov 27, 2025

Continuous non-invasive measurement of tidal volume and minute ventilation using a smart nasal cannula

Dogan AB, Patel N, Gottschalk C, Blankenship RL, Young VK, Wicks A, et al
Critical CareAdult

Study type: Observational study: it can show a link, but not prove that one thing causes another.

Read the full paper free at PubMed Central ↗

Abstract

The Flow Regulated Nasal Delivery System (FRNDS) is a novel smart nasal cannula platform developed to enable noninvasive, continuous monitoring of tidal volume (VT) and detection of abnormal breathing events. Combined with peripheral capillary oxygen saturation (SpO₂) data modulation of oxygen flow can be accomplished. In a cohort of 57 adults, FRNDS-derived VT and minute ventilation were compared to reference measurements from respiratory inductance plethysmography (RIP) across a range of oxygen flow rates. While the cannula system tended to overestimate inspiratory and underestimate expiratory VT-especially at higher flow rates-these errors were substantially reduced by applying machine learning regression models trained on anatomical and physiological features, achieving strong agreement with RIP averaging a 53.5 mL error, 78.2 mL root-mean error, and is within ~ 11% of the actual value. The system also demonstrated robust performance in classifying clinically relevant breathing patterns, including apneic spells and mouth breathing, suggesting utility for real-time respiratory surveillance and sleep apnea detection. These results support FRNDS as a promising, adaptable solution for individualized respiratory monitoring in both clinical and home environments.

Subject terms

Humans; Cannula; Tidal Volume; Male; Female; Adult; Middle Aged; Monitoring, Physiologic; Plethysmography; Aged; Oxygen Saturation; Cohort Studies

DOI: 10.1186/s13054-025-05757-1 · PubMed: 41310796

Headline, design, result, key points, caveat and citation as plain text.

Cite this article

Dogan AB, Patel N, Gottschalk C, et al. Continuous non-invasive measurement of tidal volume and minute ventilation using a smart nasal cannula. Crit Care. 2025;29(1):506. doi:10.1186/s13054-025-05757-1

RIS (EndNote, Zotero, Mendeley)BibTeX

Spotted a mistake? Suggest a correction

Top Critical Care studies every Monday

The week’s key trials and guidelines, free to read, in one short email. No account; unsubscribe in one click.

Choose specialties and topics

Similar articles

Randomized trial n = 288 · Resuscitation · Jul 22, 2021 · Pediatric · Critical Care, Pediatrics

Respiratory function monitoring did not increase target tidal volume inflations in very preterm infants

Meta-analysis n = 2,711 · Crit Care · Aug 31, 2024 · Critical Care, Cardiology

Pulse pressure and stroke volume variation outperform central venous pressure in predicting fluid response

Randomized trial n = 751 · AJRCCM · Apr 2, 2021 · Pulmonology, Critical Care

Protocolized postextubation respiratory support did not reduce reintubation rates in medical ICU patients

Systematic review · Crit Care · Feb 28, 2023 · Critical Care

High flow settings in nasal cannula therapy affect oxygenation, carbon dioxide washout, and patient comfort

Meta-analysis n = 6,806 · Intensive Care Med · Nov 25, 2021 · Critical Care, Pulmonology

Noninvasive respiratory support reduces reintubation rates compared to conventional oxygen therapy after extubation