Subcutaneous and transvenous defibrillators showed no significant difference in total device-related complications at 8-year follow-up
Device-Related Complications in Transvenous Versus Subcutaneous Defibrillator Therapy During Long-Term Follow-Up: The PRAETORIAN-XL Trial
Why it matters: It was unknown if the subcutaneous implantable cardioverter defibrillator remains superior to the transvenous version regarding device-related complications during long-term follow-up.
- Researchers followed 426 patients with a subcutaneous implantable cardioverter defibrillator and 423 patients with a transvenous implantable cardioverter defibrillator for a median of 87.5 months.
- All device-related complications were not significantly different between the two groups (P=0.15), though transvenous patients had more major complications (P=0.03) and lead-related complications (P<0.001).
- An as-treated analysis showed significantly more complications in patients with a transvenous device compared with a subcutaneous device (hazard ratio, 0.64 [95% CI, 0.41-0.99]; P=0.047).
Headline and key points written with AI from the paper’s abstract, with every number checked against it. Read the paper before relying on them.
Was this summary helpful?
Subject terms
Cite this article
Olde Nordkamp LRA, de Veld JA, Ghani A, Kuschyk J, Bonnemeier H, Bode K. Device-Related Complications in Transvenous Versus Subcutaneous Defibrillator Therapy During Long-Term Follow-Up: The PRAETORIAN-XL Trial. Circulation. 2025. doi:10.1161/CIRCULATIONAHA.125.074576