Extended Use of the Control-IQ Closed-Loop Control System in Children With Type 1 Diabetes
Abstract
Objective
To further evaluate the safety and efficacy of the Control-IQ closed-loop control (CLC) system in children with type 1 diabetes. Research design and methods
After a 16-week randomized clinical trial (RCT) comparing CLC with sensor-augmented pump (SAP) therapy in 101 children 6-13 years old with type 1 diabetes, 22 participants in the SAP group initiated use of the CLC system (referred to as SAP-CLC cohort), and 78 participants in the CLC group continued use of CLC (CLC-CLC cohort) for 12 weeks. Results
In the SAP-CLC cohort, mean percentage of time in range 70-180 mg/dL (TIR) increased from 55 ± 13% using SAP during the RCT to 65 ± 10% using CLC (P < 0.001), with 36% of the cohort achieving TIR >70% plus time <54 mg/dL <1% compared with 14% when using SAP (P = 0.03). Substantial improvement in TIR was seen after the 1st day of CLC. Time <70 mg/dL decreased from 1.80% to 1.34% (P < 0.001). In the CLC-CLC cohort, mean TIR increased from 53 ± 17% prerandomization to 67 ± 10% during the RCT and remained reasonably stable at 66 ± 10% through the 12 weeks post-RCT. No episodes of diabetic ketoacidosis or severe hypoglycemia occurred in either cohort. Conclusions
This further evaluation of the Control-IQ CLC system supports the findings of the preceding RCT that use of a closed-loop system can safely improve glycemic control in children 6-13 years old with type 1 diabetes from the 1st day of use and demonstrates that these improvements can be sustained through 28 weeks of use.
Subject terms
Cite this article
Kanapka LG, Wadwa RP, Breton MD, Ruedy KJ, Ekhlaspour L, Forlenza GP. Extended Use of the Control-IQ Closed-Loop Control System in Children With Type 1 Diabetes. Diabetes Care. 2020. doi:10.2337/dc20-1729
Spotted a mistake? Suggest a correction