Photon-counting CT: image quality evaluation in patients with tibial plateau fracture treated with metallic osteosynthesis material
Abstract
Objective
The aim was to compare the image quality of photon-counting detector CT (PCD-CT) and energy-integrating detector CT (EID-CT) in patients with tibial plateau fractures treated with metallic osteosynthesis material, and to identify optimal reconstruction parameters for PCD-CT. Materials and methods
After ethical approval, twelve patients underwent PCD-CT and EID-CT scans. Images were reconstructed using bone and soft-tissue kernels with metal artifact reduction (iMAR). PCD-CT virtual monoenergetic images (VMI) at 70, 110, and 150 keV were generated. Five radiologists assessed metal artifact severity and bone and soft-tissue visualization using a 7-point Likert scale. Visual grading characteristics analysis was performed. Noise levels were quantified and compared using Wilcoxon's signed-rank test. Results
EID-CT was rated superior in reducing metal-artifact streaks (AUC: 0.10-0.21). No significant difference was found between EID-CT iMAR and PCD-CT VMI at 110 keV and 150 keV (AUC: 0.40-0.49) concerning the metal-bone interface. An ultra-high-resolution PCD-CT kernel outperformed its EID-CT counterpart in bone visualization, with AUC values of 0.67-0.92 across all bone criteria, including those incorporating artifact-affected regions. PCD-CT showed lower noise. Conclusion
Metal artifact reduction was superior in EID-CT iMAR images compared to PCD-CT iMAR and VMI 110/150 keV, except at the metal-bone interface, where EID-CT iMAR images and PCD-CT VMI 110/150 keV performed comparably. The ultra-high-resolution PCD-CT kernel provided the best bone visualization, even when artifact-affected areas were included. Noise levels were lower in PCD-CT. This is the first in vivo comparison of photon-counting and energy-integrating CT for postoperative knee imaging with metallic osteosynthesis material. Relevance statement
These findings highlight the need for improved metal artifact reduction in PCD‑CT, while demonstrating superior bone visualization, and support a complementary interpretation strategy in postoperative knees using high‑energy VMI and MAR‑corrected reconstructions. Key points
Photon-counting CT demonstrated excellent bone visualization. Conventional EID-CT achieved better metal artifact reduction than photon-counting CT. The metal-bone interface was rated similarly for energy-integrating CT and photon-counting CT virtual monoenergetic images. These findings indicate that further improvements in metal artifact reduction algorithms for PCD-CT are warranted.